科聯電子提供多路溫度測試儀、積分球、爐溫跟蹤儀、變頻電源等產品。
咨詢熱線
0769-33213581
東莞市科聯電子有限公司
電話: 0769-33213581
傳 真:86 769 23033419
地 址:中國 廣東 東莞市 南城區莞太路233號勝華產業園3棟2樓
郵 編:523000
2018-06-11 來源:科聯電子
積分球LED光電色檢測操作規程
Operation procedure for the detection of the photoelectric color of the integral sphere LED
第一次使用儀器
The first use of the instrument
1)把軟件打開, 保證0.3米積分球和1.5米積分球是關閉狀態,積分球內不能放任何光源。
1) to open the software, to ensure that the 0.3 meters and 1.5 meters of integrated ball is off the ball is closed, integral ball can not put any light source.
2)點擊軟件上面清零 ok 等待此任務即可
2) click on the software to clear the OK waiting for this task
開啟檢測設備
Opening test equipment
開啟配電機柜總電源,開關位置為:機柜中間面板的左側船型開關;
Turn on the power switch of the power distribution cabinet and the switch position is: the left side of the middle panel of the cabinet;
開啟機柜內部各個儀表的電源(功率計、穩壓變頻電源) ;
Power supply (power meter, voltage regulated power supply) for various instruments in the cabinet;
開啟直流電源;
Open DC power supply;
1.5米積分球的校準測量步驟
1.5 meter measuring steps
運行桌面1.5米球的測試軟件(確認桌面0.3米球的軟件已經關閉);
Test software running 1.5 meters of the table, the software has been closed down for 0.3 meters;
在1. 5米球內裝入50W標準光源;
50W standard light source in the 1.5 meter ball;
切換機柜右側的船型開關到1.5米積分球;
Switch on the right side of the switch to the 1.5 meter integral ball ;
保持機柜中間面板的中間轉換開關到“DC/直流”;
To maintain the intermediate panel of the intermediate switch to the "DC/ dc";
在軟件上點進入界面把電壓 電流 流明 色溫 設置好,數據已標準燈為準。
On the software point of entry interface, the voltage and current lumen color temperature is set up, and the data has been standard lamp.
數據設置好后,點亮標準光源(標準燈需要預熱5分鐘時間)
After the data is set up, light up the standard light source (the standard lamp needs to warm for 5 minutes).
標準光源點亮后,直接點校準,預熱時間過后會出現一個山峰曲線圖譜后,點擊保存退出
After the standard light source is illuminated, it will be calibrated directly. After the preheating time, there will be a map of the peak curve. Click Save and exit
等待校準完成后再點擊“開始測試”;查看對比結果。
Click the "Instrument calibration" button, select "spectral calibration", select "color temperature flux calibration"; enter the standard color temperature, standard flux; click on the "Calibration" and then wait for the calibration is complete click on "Start Test";. For a comparison of the results.
1.5米積分球校準定標完成。
1.5 meters integral sphere calibration is completed.
等待2分鐘,確認標準光源已經冷卻,取下標準光源,放入燈盒中;開始測試貴公司的產品
Wait for 2 minutes, confirm the standard light source has cooled, take the standard light source, into the lamp box;
如果產品是交流的話關閉穩壓直流,開啟變頻測試電源。
Close the regulator DC, open the power supply of variable frequency test.
轉換機柜中間面板的中間轉換開關到“AC/交流”
Conversion switch to "AC/""
點亮被測燈具,點擊“開始測試”
Light up the lamp, click "start testing"
0.3米積分球的校準測量步驟
0.3 meter measuring steps
運行桌面0.3米球的測試軟件(確認桌面1.5米球的軟件已經關閉);
Test software running 0.3 meters of the table, the software has been closed down for 1.5 meters;
在0.3米球內裝入10W標準光源;小球軟件
Load 10W standard light source within 0.3 m ball;
切換機柜右側的船型開關到0.3米積分球
Switch on the right side of the switch to the 0.3 meter integral ball
保持機柜中間面板的中間轉換開關到“DC/直流”
To maintain the intermediate panel of the intermediate switch to the "DC/ dc"
點亮標準光源
Light standard light source
標準光源點亮后,穩定5分鐘; 軟件操作跟大球一樣。
Standard light source, stable for 2 minutes;
點擊開始測試,查看對比結果。
Click to start the test to see the contrast results.
點擊“儀器校準”按鈕,選擇“光譜校準”,選擇“色溫光通量校準”;輸入標準色溫,標準光通量;點擊“校準”。等待校準完成后再點擊“開始測試”;查看對比結果。
Click the "Instrument calibration" button, select "spectral calibration", select "color temperature flux calibration"; enter the standard color temperature, standard flux; click on the "Calibration" and then wait for the calibration is complete click on "Start Test";. For a comparison of the results.
0.3米積分球校準定標完成。
0.3 meters integral sphere calibration is completed.
等待2分鐘,確認標準光源已經冷卻,取下標準光源,放入燈盒中;
Wait for 2 minutes, confirm the standard light source has cooled, take the standard light source, into the lamp box;
將被測LED放入相應的專用LED夾具,然后放入到0.3米積分球側面開口處,固定好,夾具有(鹵鎢燈,大功率1W,3528,5050,3014等);
Will be measured LED into the corresponding special LED fixture, and then put into the 0.3 meters of the integral ball side openings, fixed, with (halogen tungsten lamp, high power 1W, 352850503014, etc.);
調節直流恒流電源,點亮被測LED;
Adjust the DC constant current power supply, light is measured LED;
點擊“開始”按鈕,等待測量完成(光譜及電參數會自動完成測試);
Click "start" button and wait for the measurement to finish (the spectrum and the electrical parameters will be automatically finished testing);
點擊“保存”按鈕,記錄下原始數據文件;
Click the Save button to record the raw data file;
導出PDF測試報告,過程:點擊“文件”—“導出”—“PDF”;
Export PDF test report, process: click the "file" - "export" - "PDF";
重復13)到19)測量下一個被測燈;
Repeat 13) to measure the next measured light by 19;